Page 18 - Healthcare
P. 18

/ Point-of-use filters


                               CONTROLLING WATERBORNE                    CURRENT GUIDANCE
                               HEALTH RISKS



                               Water, essential for hygiene, can also be a source   HSE guidance HSG274 Part 2 (2014) and DHSC
                               of infection if its quality is not controlled.   Health Technical Memorandum (HTM) 04-01
                               The proliferation of bacteria (Legionella spp.,   Part B (2016) provide guidance for the control
                               Pseudomonas aeruginosa, etc.) in the water supply   of Legionella in hot and cold water systems and
                               or in mixers and taps may cause serious infections,   on providing safe water in healthcare premises
                               especially for vulnerable people.         respectively
                                                                         Point-of-use filters prevent the discharge of planktonic
                               Water systems in all buildings open to the public   Legionella from the tap and shower outlets and may be
          BIOFIL filter for    should be monitored for Legionella (e.g. hotels   used to provide water free of P. aeruginosa.
          mixers and taps      and holiday accommodation, campsites, detention   They should be used primarily as a temporary
                               centres, etc.) and not just healthcare facilities.   measure until a permanent, safe engineering solution
                               The HSE Approved Code of Practice and Guidance   is developed, although long-term use of such filters
                               (ACOP) document L8 states that duty holders should   may be needed in some healthcare situations.
                               identify and assess sources of risk of exposure    They may also be considered where a high level of
                               to Legionella; put in place precautions to prevent   disinfection of water systems may dislodge biofilm.
                               or control that risk; and monitor these measures to   Point-of-use micro-filters with a 0.2µm membrane
                               ensure that they remain effective. This applies in all   constitute a means for removing Legionella and
                               circumstances where the Health and Safety at Work   Pseudomonas aeruginosa in water drawn off from
                               Act 1974 applies.                         the system.
                               The prevention of health risks associated with   HTM 04-01 Part C (2016) provides advice
                               pathogenic bacteria in water systems should be a   on managing Pseudomonas aeruginosa in
                               constant concern for public buildings, and in particular   augmented care units
                               Healthcare Facilities (see the current guidance   For direct contact with augmented care patients,
          BIOFIL spout filter   opposite).                               water of a known satisfactory quality should be used,
          for BIOCLIP mixers and taps                                    that is, either: water where testing has shown absence
                                                                         of P. aeruginosa; or water supplied through
                               The BIOFIL range of point-of-use filters provides a
                               preventative or curative solution to secure the water   a POU filter; or sterile water (for example, for skin
                               quality at the point-of-use and protect the health of   contact for babies in neonatal intensive care units).
                               the user.
                                                                         HSG274 Part 2 states that monitoring
                                                                         for Legionella should be carried out:
                                                                         - when water is treated with biocides and when
                                                                            distribution temperatures are reduced from HSE
                                                                            guidelines.
                                                                         - where water treatment regime control levels are not
                                                                            being consistently achieved.
                                                                         - where there is a high-risk of contamination e.g.
                                                                            hospitals and care homes.
                                                                         - where an outbreak of legionellosis has been
          BIOFIL shower head filter                                         identified.
                                                                         When a sample identifies a Legionella count above
                                                                         100 CFU/Litre (Colony Forming Unit) an immediate
                                                                         review of the control measures and risk assessment
                                                                         should be carried out to identify any remedial action
                                                                         required. Disinfection of the system should also be
                                                                         considered.
                                                                         Point-of-use micro-filters are a means to restore
                                                                         the bacteriological quality of water at the point-of-
                                                                         use (anti-Legionella and all-germ tap filters and
                                                                         shower heads, etc.).





          BIOFIL filter
          for wall-mounted showers














          16  Point-of-use filters